TA202
www.baumer.com 15
9. Programming plan
line 11-16 status for lines 1-6
0 * Alter parameter
1 Lock of delete resp.enter
2 Skip parameter
line 21 Calcuating functions
0 * Relation tacho F1 : tacho F2
1 difference tacho F1 - tacho F2
2 stretching / compression
3 Flow time (length)
4 Time measurement - Start/Stopp
5 Time measurement – period
6 Time measurement – pulse duration
7 Pulse rate measurement
line 22 Inversion of calculating function
0 * Calculating function as above
1 inverted F1->F2 F2->F1
line 23 Count mode tacho for F1 and F2
0 * F1 Track A with Up/Down F2 40 kHz
1 F1 Track A 90° B F2 40 kHz
2 F1 Track A with Up/Down F2 25 Hz
3 F1 Track A 90° B F2 25 Hz
line 24 Input logic (for F1 and F2)
0 * PNP Trigger level approx. 11V
1 NPN Trigger level approx. 11V
2 PNP Trigger level approx. 5V
3 NPN Trigger level approx. 5V
4 PNP Trigger level approx. 2,5V
5 NPN Trigger level approx. 2,5V
line 25 Evaluation tacho F1
0.0001 Min.
1.0000 * Default
9999.99 Max.
line 26 Evaluation tacho F2
0.0001 Min.
1.0000 * Default
9999.99 Max.
line 27 Factor for calculation function F3
0.0001 Min.
1.0000 * Default
9999.99 Max.
line 28 minimum update time for F1-F3
0 Maximum update time (approx. 0,5s)
1 * 1 second
2 2 Seconds
3 3 Seconds
4 5 Seconds
5 10 Seconds
6 20 Seconds
7 30 Seconds
8 60 Seconds
line 29 Time out for F1-F3
0 * 1 second (0,1S)
1 2 Seconds (0,2s)
2 3 Seconds (0,3s)
3 5 Seconds (0,5s)
4 10 Seconds (1s)
5 20 Seconds (2s)
6 30 Seconds (3s)
7 60 Seconds (6s)
8 Time-out function out of order (8s)
9 as 8 but store (10s)
line 30 Time unit tacho F1
0 * 1/min
1 1/s
2 1/h
line 31 Time unit tacho F2
0 * 1/min
1 1/s
2 1/h
line 32 Assignment of dragging value
0 * F1
1 F2
2 F3
line 33 Assignment output 1
0 * Upper Limit value tacho F1
1 Lower Limit value tacho F1
2 Upper Limit value tacho F2
3 Lower Limit value tacho F2
4 Upper Limit value calculating function F3
5 Lower Limit value calculating function F3
line 34 Assignment output 2
0 * Upper Limit value tacho F1
1 Lower Limit value tacho F1
2 Upper Limit value tacho F2
3 Lower Limit value tacho F2
4 Upper Limit value calculating function F3
5 Lower Limit value calculating function F3
line 35 Output logic for digital output
0 * Both outputs as normally closed
1 Out 1 open, Out 2 closed
2 Out 1 closed, Out 2 open
3 Both outputs as normally open
line 36 Decimal point for tacho F1
0 * No decimal point
1 00000.0
2 0000.00
3 000.000
line 37 Decimal point for tacho F2
0 * No decimal point
1 00000.0
2 0000.00
3 000.000
line 38 Decimal point for calculating function F3
Calc. func. 0-2,7 Calc.func. 3-6
0 * No decimal point 59.59.99 min
1 00000.0 99.59.59 h
2 0000.00 "
3 000.000 "
line 39 Basic display selection
0 * No switchover to basic display
1 F1
2 F2
3 F3
4 SZ
5 P1
6 P2
line 40 Code
0000 * Code not active
:
9999
line 41 Function key assignment
0 * Function key not assigned
:
6 Maximum value (line 6)
line 42 Output reaction of lower limit value
1 * with contact lock
0 W/o contact lock
line 51 Baudrate
0 * 4800 Baud
1 2400 Baud
2 1200 Baud
3 600 Baud
line 52 Parity
0 * Even Parity
1 Odd Parity
2 No Parity
Line 53 Stop bits
0 * 1 Stop bit
1 2 Stop bits
Line 54 Device address
00 * Minimum value
:
99 Maximum value
Kommentare zu diesen Handbüchern